Pirani gauge

From Infogalactic: the planetary knowledge core
Jump to: navigation, search
File:Pirani sensor.jpg
Pirani probe with 64 mm vacuum flange

The Pirani gauge is a robust thermal conductivity gauge used for the measurement of the pressures in vacuum systems.[1] It was invented in 1906 by Marcello Pirani.[2]

Structure

The Pirani gauge consists of a metal filament (usually platinum) suspended in a tube which is connected to the system whose vacuum is to be measured. Connection is usually made either by a ground glass joint or a flanged metal connector, sealed with an o-ring. The filament is connected to an electrical circuit from which, after calibration, a pressure reading may be taken.

Mode of operation

File:Blockbild-Pirani-Vakuumeter.png
Block diagram of Pirani gauge
File:Nomogramm pirani-Vakuummeter.jpg
Curves to convert air readings to other gases

A heated metal wire (also called a filament) suspended in a gas will lose heat to the gas as its molecules collide with the wire and remove heat. If the gas pressure is reduced the number of molecules present will fall proportionately and the wire will lose heat more slowly. Measuring the heat loss is an indirect indication of pressure.

The electrical resistance of a wire varies with its temperature, so the resistance indicates the temperature of wire. In many systems, the wire is maintained at a constant resistance R by controlling the current I through the wire. The resistance can be set using a bridge circuit. The power delivered to the wire is I2R, and the same power is transferred to the gas.[3] The current required to achieve this balance is therefore a measure of the vacuum.

The gauge may be used for pressures between 0.5 Torr to 10−4 Torr. The thermal conductivity and heat capacity of the gas may affect the readout from the meter, and therefore the apparatus may need calibrating before accurate readings are obtainable. For lower pressure measurement other instruments such as a Penning gauge are used.

Pulsed Pirani gauge

A special form of the Pirani gauge is the pulsed Pirani vacuum gauge where the filament is not operated at a constant temperature, but is cyclically heated up to a certain temperature threshold by an increasing voltage ramp. When the threshold is reached, the heating voltage is switched off and the filament cools down again. The required heat-up time is used as a measure of pressure.

For adequately low pressure the following relation for supplied heating power and filament temperature T(t) applies:[4]

P_{\text{el}} = C_1 \lambda_{\text{gas}}(T(t) - T_a) + C_2\lambda_{\text{fil}}(T(t) - T_a) + A_{\text{fil}} \epsilon \sigma(T(t)^4 - T^4_a) + c_{\text{fil}}m_{\text{fil}} \frac{\mathrm{d}T}{\mathrm{d}t} ,

where c_{\text{fil}} is the heating capacity of the filament, m_{\text{fil}} is the mass of the filament and C_1 and C_2 are constants.

Advantages and disadvantages of the pulsed gauge

Advantages
  • Significantly better resolution in the range above 75 Torr. [5]
  • The power consumption is drastically reduced compared to continuously operated Pirani gauges.
  • The gauge's thermal influence on the real measurement is lowered considerably due to the low temperature threshold of 80°C and the ramp heating in pulsed mode.
  • The pulsed mode allows for efficient implementation of modern microprocessor technology.
Disadvantages
  • Increased calibration effort
  • Longer heat-up phase

Alternative

An alternative to the Pirani gauge is the thermocouple gauge, which works on the same principle of detecting thermal conductivity of the gas by a change in temperature. In the thermocouple gauge, the temperature is sensed by a thermocouple rather than by the change in resistance of the heated wire.

References

  1. Lua error in package.lua at line 80: module 'strict' not found.
  2. Lua error in package.lua at line 80: module 'strict' not found.
  3. Ignoring the power lost by the electrical connections and the wire supports.
  4. DE 10115715, Plöchinger, Heinz, "Sensor and Method for Detecting Measurement Variables and Physical Parameters", published 2001-03-30, issued 2002-10-17 , also description
  5. Lua error in package.lua at line 80: module 'strict' not found.

External links